Service Issues                                              Possible Cause                                                                Suggestions

Scale displays “Err” after initial plug in

1. Platform cable not connected to console

2. Cable from platform damaged

3. Load cell problem

1. Check transducer cable connection

2. Check for cut or damaged wires

3. Call for service

Scale won’t turn on

1. Disposable alkaline batteries are weak

2. Bad Keypad

 

1. Replace with new batteries

2. Call for repair

Scale displays "ERR-2" or "ERR-3"

1. Capacity of the scale has been exceeded

 

1. Do not place more than the rated capacity on platform

Scale displays "ERR-4"

1. Insufficient weight on platform to calibrate

Use at least 100 lbs. to calibrate

Scale is drifting

1. Move to dry location

2. Firmly plug in all cables

Erratic weight readings on platform

1. Scale touching wall or desk

2. Leveling feet not screwed tightly

3. Scale on deep plush carpet

1. Move away from wall or desk

2. Tightly screw in all four leveling feet

3. Move to harder surface

Incorrect weights

1. Scale is out of calibration

1. Follow calibration instructions

LB/KG button does not function

1. Button lock has been activated

2. Height has been set in BMI mode before user stands on platform

1. See manual

2. Press "ON/ZERO" to clear height or step on scale platform

 

1. Moisture in platform

2. Loose cable connection